Tesla | 0.3T | 0.35T | 0.4T | 0.5T | |
---|---|---|---|---|---|
Operation Frequency (*1) | 12.77MHz | 14.90MHz | 17.00MHz | 21.29MHz | |
Frequency Range | < +/-200kHz | ||||
Peak Output Power | 4ch | 1.25kW x 4 (Total 5kW) | |||
2ch | 2.5kW x 2 (Total 5kW) | ||||
1ch | 5kW | ||||
Cooling | Air | ||||
RF Input | 0dBm x1 | ||||
Impedance | 50ohm | ||||
VSWR | 4.0:1 or less | ||||
Gain | 61dB(4ch)/ 64dB(2ch)/ 67dB(1ch) | ||||
Average Power | 250W | ||||
Max Gate On Time | 60msec | ||||
Max Gate On Duty | 50% maximum for low output power
15% maximum @ Rated output power |
||||
Gain Linearity | ±1.0dB or less(50Ω) | ||||
Phase Linearity(between ch1 and ch2) | 0 ± 5 degrees | ||||
Droop (Pulse Wide = 3.5msec) | - | ||||
Harmonics | -30 dBc or less (50Ω) | ||||
Output Noise Power <Blanked>
(Center Frequency +/-200kHz) |
-160dBm/Hz or less | ||||
Interface | RS422 conforming
TTL(Gate Signal) |
||||
Main Voltage | single phase AC 220V ±10% | ||||
Dimension | Height | 177mm | |||
Depth | 567mm | ||||
Width | 478mm |
(*1) Operation Frequency will be customized by MRI maker's request.