| Tesla | 0.3T | 0.35T | 0.4T | 0.5T | |
|---|---|---|---|---|---|
| Operation Frequency (*1) | 12.77MHz | 14.90MHz | 17.00MHz | 21.29MHz | |
| Frequency Range | < +/-200kHz | ||||
| Peak Output Power | 4ch | 1.25kW x 4 (Total 5kW) | |||
| 2ch | 2.5kW x 2 (Total 5kW) | ||||
| 1ch | 5kW | ||||
| Cooling | Air | ||||
| RF Input | 0dBm x1 | ||||
| Impedance | 50ohm | ||||
| VSWR | 4.0:1 or less | ||||
| Gain | 61dB(4ch)/ 64dB(2ch)/ 67dB(1ch) | ||||
| Average Power | 250W | ||||
| Max Gate On Time | 60msec | ||||
| Max Gate On Duty | 50% maximum for low output power
15% maximum @ Rated output power |
||||
| Gain Linearity | ±1.0dB or less(50Ω) | ||||
| Phase Linearity(between ch1 and ch2) | 0 ± 5 degrees | ||||
| Droop (Pulse Wide = 3.5msec) | - | ||||
| Harmonics | -30 dBc or less (50Ω) | ||||
| Output Noise Power <Blanked>
(Center Frequency +/-200kHz) |
-160dBm/Hz or less | ||||
| Interface | RS422 conforming
TTL(Gate Signal) |
||||
| Main Voltage | single phase AC 220V ±10% | ||||
| Dimension | Height | 177mm | |||
| Depth | 567mm | ||||
| Width | 478mm | ||||
(*1) Operation Frequency will be customized by MRI maker's request.
For product details, contact the sales manager CMSC-PS_components@medical.canon
© CANON MEDICAL SYSTEMS DO BRASIL LTDA.